FDA Adverse Event Injury Summary report: N

CAPSUREFIX

MDR report key: 3991220 · Received August 8, 2014

Report

Report Number
2649622-2014-08236
Event Type
Injury
Date Received
August 8, 2014
Date of Event
May 21, 2014
Report Date
May 21, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE LEAD WAS RETURNED IN SEGMENTS, ANALYZED, AND NO ANOMALIES WERE FOUND. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED APPARENT EXPLANT DAMAGE. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD WAS NOT CAPTURING. DURING AN EXTRACTION ATTEMPT THE LEAD WAS ADHERED TO ANOTHER LEAD. THE LEAD WAS INACTIVATED AND WAS SUBSEQUENTLY LASER EXTRACTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
468015 CAPSUREFIX ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5568-53

Patients

Seq Age Sex Outcome Treatment
1 00078 YR Hospitalization| R VEDR01 IPG, 5076 LEAD