FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 3991216 · Received August 8, 2014

Report

Report Number
2649622-2014-08239
Event Type
Injury
Date Received
August 8, 2014
Date of Event
April 10, 2014
Report Date
May 2, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Removal / Correction Number
Z-0475-2011
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: D284DRG ICD, IMPLANTED: (B)(6) 2009. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED OVERSENSING DUE TO NON-PHYSIOLOGIC SIGNALS/SIC (SENSING INTEGRITY COUNTER). THERE WERE 5 NON-SUSTAINED TACHYCARDIA EPISODES WITH CYCLE LENGTH INTERVALS LESS THAN 220 MS BETWEEN (B)(6) 2014 AND (B)(6) 2014. LEAD FAILURE PREDICTOR TRIGGERED ON (B)(6) 2014 FOR VENTRICULAR-SICS AND NON-SUSTAINED TACHYCARDIA EPISODES. THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION. BASED ON THE INFORMATION RECEIVED AND WITHOUT THE RETURN OF THE PRODUCT, IT COULD NOT DETERMINE THIS DEVICE PERFORMED AS DESCRIBED IN THE FIELD ACTION.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD TRIGGERED A LEAD INTEGRITY ALERT (LIA) DUE TO A HIGH NUMBER OF SHORT INTERVAL COUNTS (SIC). THERE WAS ALSO INTERMITTENT OVERSENSING AND NOISE. LEAD TESTING AND POCKET MANIPULATION WERE PERFORMED. THE LEAD WAS REPROGRAMMED AND REMAINS IN USE WITH MONTHLY MONITORING. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
467353 SPRINT QUATTRO SECURE DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 694765

Patients

Seq Age Sex Outcome Treatment
1 00071 YR Required Intervention 5076-58 LEAD