FDA Adverse Event
Injury
Summary report: N
CAPSUREFIX MRI SURESCAN
MDR report key: 3991209
·
Received August 8, 2014
Report
- Report Number
- 2649622-2014-08243
- Event Type
- Injury
- Date Received
- August 8, 2014
- Date of Event
- May 5, 2014
- Report Date
- May 5, 2014
- Manufacturer
- MPRI
- Product Code
- NVN
- PMA / PMN Number
- P090013
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED AND ANALYSIS WAS PERFORMED AND NO ANOMALIES WERE FOUND. NO TISSUE IN THE HELIX AT TIME OF ANALYSIS. CONCOMITANT PRODUCT: 5086MRI52 LEAD IMPLANTED: (B)(6) 2014. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD WAS FOUND DISLODGED POST IMPLANT. IT WAS NOTED THERE WAS TISSUE IN THE HELIX. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 472513 | CAPSUREFIX MRI SURESCAN | DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE | NVN | MPRI | 5086MRI45 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00074 YR | Hospitalization| R | RVDR01 IPG |