FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 3991199 · Received August 8, 2014

Report

Report Number
2649622-2014-08244
Event Type
Injury
Date Received
August 8, 2014
Date of Event
November 18, 2013
Report Date
May 21, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYZED. NO ANOMALIES WERE FOUND. THE INNER AND OUTER INSULATION OF THE LEAD WAS EXTRINSICALLY BREACHED DUE TO A CUT. DUE TO THE LENGTH OF IMPLANT AND THE MINIMAL AMOUNT OF BLOOD INGRESS IT IS LIKELY THAT THE CUT OCCURRED DURING THE EXPLANT. ALL ELECTRICAL TESTING WAS WITHIN SPECIFIED PARAMETERS. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED APPARENT EXPLANT DAMAGE. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT NOISE WAS PRESENT ON THE RIGHT VENTRICULAR LEAD. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
467082 SPRINT QUATTRO SECURE DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 694765

Patients

Seq Age Sex Outcome Treatment
1 00067 YR Hospitalization| R ST. JUDE MEDICAL ICD