FDA Adverse Event Injury Summary report: N

CONSULTA CRT-D

MDR report key: 3991106 · Received August 8, 2014

Report

Report Number
3004209178-2014-14347
Event Type
Injury
Date Received
August 8, 2014
Date of Event
June 24, 2014
Report Date
June 24, 2014
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
NIK
PMA / PMN Number
P010031
Removal / Correction Number
Z-0111-2011
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: 6725 CRDM ADAPTOR, IMPLANTED (B)(6) 2010; 419488 LEAD, IMPLANTED (B)(6) 2010. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS INDICATED HIGH IMPEDANCE IN THE BATTERY. ANALYSIS OF THE DEVICE MEMORY FOUND THE ERI (ELECTIVE REPLACEMENT INDICATOR) IS THE RESULT OF HIGH INTERNAL BATTERY RESISTANCE. THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION, BUT RETURNED PRODUCT TESTING FOUND THE DEVICE DID NOT PERFORM AS DESCRIBED IN THE FIELD ACTION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS EXPLANTED AND REPLACED AS IT COULD NOT BE INTERROGATED. IT WAS ALSO REPORTED THAT THE LEFT VENTRICULAR (LV) LEAD HAD EXHIBITED CHRONIC HIGH THRESHOLD MEASUREMENTS. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
471856 CONSULTA CRT-D DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO NIK MEDTRONIC MED REL MEDTRONIC PUERTO RICO D224TRK

Patients

Seq Age Sex Outcome Treatment
1 00085 YR Hospitalization| R 693565 LEAD