FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX MRI SURESCAN

MDR report key: 3991023 · Received August 8, 2014

Report

Report Number
2649622-2014-08330
Event Type
Malfunction
Date Received
August 8, 2014
Date of Event
June 12, 2014
Report Date
June 12, 2014
Manufacturer
MPRI
Product Code
NVN
PMA / PMN Number
P090013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYZED. ANALYSIS WAS PERFORMED AND NO ANOMALIES WERE FOUND. THE DISTAL LV (LOW VOLTAGE) ELECTRODE OF THE LEAD WAS COVERED IN BLOOD AND BODY TISSUE/FIBROTIC GROWTH. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT AND STRETCHING. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN IMPLANT ATTEMPT, THE PHYSICIAN HAD TO REPOSITION THE ATRIAL LEAD SEVERAL TIMES DUE TO PATIENT ANATOMY. LAST TIME DURING REPOSITIONING THE HELIX WOULD NOT RETRACT BACK AFTER OVER 50 TURNS. ANOTHER LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
471719 CAPSUREFIX MRI SURESCAN DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE NVN MPRI 5086MRI45

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R