CAPSURE Z NOVUS
Report
- Report Number
- 2649622-2014-08328
- Event Type
- Injury
- Date Received
- August 8, 2014
- Date of Event
- June 10, 2014
- Report Date
- June 16, 2014
- Manufacturer
- MPRI
- Product Code
- DTB
- PMA / PMN Number
- P850089
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: 5554, IMPLANTABLE PACING LEAD, IMPLANTED: (B)(6) 2009. (B)(4).
PRODUCT EVENT SUMMARY: THE LEAD WAS RETURNED TO THE MANUFACTURER AND ANALYZED. THE OUTER INSULATION OF THE LEAD DEVELOPED A BREACH DUE TO A DEPRESSION WHILE IN VIVO, THE OUTER INSULATION OF THE LEAD DEVELOPED A COSMETIC DEPRESSION WHILE IN VIVO, AND THE OUTER INSULATION OF THE LEAD WAS OBSERVED TO HAVE BLOOD INGRESSION.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD WAS NOT CAPTURING. THE RV LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 470999 | CAPSURE Z NOVUS | ELECTRODE, PACEMAKER, PERMANENT | DTB | MPRI | 5054-52 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00089 YR | Hospitalization| R | P1501DR IPG |