FDA Adverse Event Injury Summary report: N

CAPSUREFIX MRI SURESCAN

MDR report key: 3991014 · Received August 8, 2014

Report

Report Number
2649622-2014-08360
Event Type
Injury
Date Received
August 8, 2014
Date of Event
June 16, 2014
Report Date
June 16, 2014
Manufacturer
MPRI
Product Code
NVN
PMA / PMN Number
P090013
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYZED. ANALYSIS REVEALED THE INNER INSULATION HAD A DEPRESSION BREACH.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYZED. THE PROXIMAL CONDUCTOR OF THE LEAD BECAME EXTRINSICALLY DISTORTED DUE TO KINKING/BUCKLING. THE DISTAL CONDUCTOR WAS EXTRINSICALLY DISTORTED DUE TO KINKING/BUCKLING . THE ANALYST COMMENTED, ELECTRICAL CROSS CONTINUITY TEST RESULT WAS FAILED. PERFORMED DESTRUCTIVE ANALYSIS AND FOUND INNER INSULATION BREACHED IN -VIVO/ DEPRESSION. INSULATION BREACH DID CONTRIBUTE TO THE ELECTRICAL COMPLAINTS. CONCOMITANT MEDICAL PRODUCTS: 5086MRI58, LEAD, (B)(6) 2011. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THE ATRIAL LEAD HAD EXIT BLOCK, HIGH THRESHOLDS, AND INTERMITTENT PACING. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
470820 CAPSUREFIX MRI SURESCAN DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE NVN MPRI 5086MRI52

Patients

Seq Age Sex Outcome Treatment
1 00057 YR Hospitalization| R