CAPSUREFIX MRI SURESCAN
Report
- Report Number
- 2649622-2014-08360
- Event Type
- Injury
- Date Received
- August 8, 2014
- Date of Event
- June 16, 2014
- Report Date
- June 16, 2014
- Manufacturer
- MPRI
- Product Code
- NVN
- PMA / PMN Number
- P090013
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYZED. ANALYSIS REVEALED THE INNER INSULATION HAD A DEPRESSION BREACH.
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYZED. THE PROXIMAL CONDUCTOR OF THE LEAD BECAME EXTRINSICALLY DISTORTED DUE TO KINKING/BUCKLING. THE DISTAL CONDUCTOR WAS EXTRINSICALLY DISTORTED DUE TO KINKING/BUCKLING . THE ANALYST COMMENTED, ELECTRICAL CROSS CONTINUITY TEST RESULT WAS FAILED. PERFORMED DESTRUCTIVE ANALYSIS AND FOUND INNER INSULATION BREACHED IN -VIVO/ DEPRESSION. INSULATION BREACH DID CONTRIBUTE TO THE ELECTRICAL COMPLAINTS. CONCOMITANT MEDICAL PRODUCTS: 5086MRI58, LEAD, (B)(6) 2011. (B)(4).
IT WAS REPORTED THE ATRIAL LEAD HAD EXIT BLOCK, HIGH THRESHOLDS, AND INTERMITTENT PACING. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 470820 | CAPSUREFIX MRI SURESCAN | DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE | NVN | MPRI | 5086MRI52 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00057 YR | Hospitalization| R |