FDA Adverse Event Injury Summary report: N

MARQUIS VR

MDR report key: 3991008 · Received August 8, 2014

Report

Report Number
9614453-2014-01735
Event Type
Injury
Date Received
August 8, 2014
Date of Event
May 13, 2014
Report Date
June 25, 2014
Manufacturer
IPG MFG SWITZERLAND
Product Code
LWS
PMA / PMN Number
P980016
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY - THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. ANALYSIS REVEALED A WRITE TO LOCKED RAM POWER ON RESET WAS RECORDED ON (B)(6) 2013.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE HAD AN ELECTRICAL RESET. THE PATIENT HAD UNDERGONE RADIATION THERAPY. THE DEVICE WAS RESET AND REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
470818 MARQUIS VR DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS IPG MFG SWITZERLAND 7230CX

Patients

Seq Age Sex Outcome Treatment
1 00068 YR Required Intervention