FDA Adverse Event
Injury
Summary report: N
MARQUIS VR
MDR report key: 3991008
·
Received August 8, 2014
Report
- Report Number
- 9614453-2014-01735
- Event Type
- Injury
- Date Received
- August 8, 2014
- Date of Event
- May 13, 2014
- Report Date
- June 25, 2014
- Manufacturer
- IPG MFG SWITZERLAND
- Product Code
- LWS
- PMA / PMN Number
- P980016
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).
Additional Manufacturer Narrative · 1
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
Additional Manufacturer Narrative · 1
PRODUCT EVENT SUMMARY - THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. ANALYSIS REVEALED A WRITE TO LOCKED RAM POWER ON RESET WAS RECORDED ON (B)(6) 2013.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE HAD AN ELECTRICAL RESET. THE PATIENT HAD UNDERGONE RADIATION THERAPY. THE DEVICE WAS RESET AND REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 470818 | MARQUIS VR | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | IPG MFG SWITZERLAND | 7230CX |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00068 YR | Required Intervention |