FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 3990979 · Received August 8, 2014

Report

Report Number
2649622-2014-08366
Event Type
Injury
Date Received
August 8, 2014
Date of Event
May 8, 2014
Report Date
May 8, 2014
Manufacturer
MPRI
Product Code
NVY
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: 407645 LEAD, IMPLANTED (B)(6) 2012. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD TRIGGERED AN ALERT FOR HIGH SUPERIOR VENA CAVA (SVC) IMPEDANCE. THE PHYSICIAN DECIDED TO PROGRAM THE SVC COIL TO OFF AND THE ALERTS FOR SVC IMPEDANCE WERE ALSO TURNED OFF. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
471519 SPRINT QUATTRO SECURE DRUG-ELUTING PERMANENT DEFIBRILLATOR ELECTRODES NVY MPRI 6947M55

Patients

Seq Age Sex Outcome Treatment
1 00040 YR Required Intervention D314DRM ICD