FDA Adverse Event Injury Summary report: N

SELECTSECURE

MDR report key: 3990951 · Received August 8, 2014

Report

Report Number
2649622-2014-08401
Event Type
Injury
Date Received
August 8, 2014
Date of Event
May 21, 2014
Report Date
May 21, 2014
Manufacturer
MPRI
Product Code
NVN
PMA / PMN Number
P030036
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCTS: (B)(4) ICD IMPLANTED: 2010 (B)(6); 7120 COMPETITOR LEAD IMPLANTED: 2010 (B)(6). (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DISTAL PORTION OF THE LEAD WAS RETURNED, ANALYZED, AND THE OUTER INSULATION OF THE LEAD DEVELOPED A BREACH DUE TO ESC (ENVIRONMENTAL STRESS CRACKING) WHILE IN VIVO. IT WAS ALSO NOTED THAT THE OUTER INSULATION OF THE LEAD DEVELOPED COSMETIC ESC (ENVIRONMENTAL STRESS CRACKING) WHILE IN VIVO AND THE OUTER INSULATION OF THE LEAD WAS OBSERVED TO HAVE BLOOD INGRESSION. THE ANALYSIS COMMENTS STATED THAT THE VISUAL ANALYSIS FOUND THE OUTER INSULATION BREACHED IN ¿VIVO/ ENVIRONMENTAL STRESS CRACKING AND THAT THE INSULATION BREACHED DID CONTRIBUTE TO THE ELECTRICAL COMPLAINT OF HIGH THRESHOLDS.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE ATRIAL LEAD HAD HIGH THRESHOLDS. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
469858 SELECTSECURE DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE NVN MPRI 383059

Patients

Seq Age Sex Outcome Treatment
1 00082 YR Hospitalization| R 419488 LEAD