FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX MRI SURESCAN

MDR report key: 3990899 · Received August 8, 2014

Report

Report Number
2649622-2014-08407
Event Type
Malfunction
Date Received
August 8, 2014
Date of Event
May 9, 2014
Report Date
May 9, 2014
Manufacturer
MPRI
Product Code
NVN
PMA / PMN Number
P090013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY : THE FULL LEAD WAS RETURNED AND ANALYZED WITH NO ANOMALIES FOUND. THERE WAS A BREACHED CUT ON THE OUTER INSULATION AND VISUAL ANALYSIS NOTED THE LEAD WAS DAMAGED AT IMPLANT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN IMPLANT PROCEDURE, THE RIGHT VENTRICULAR LEAD HAD LOW R WAVE AMPLITUDES IN VARIOUS POSITIONS. THE LEAD WAS NOT USED AND ANOTHER LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
470298 CAPSUREFIX MRI SURESCAN DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE NVN MPRI 5086MRI52

Patients

Seq Age Sex Outcome Treatment
1 00073 YR