FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 3990859 · Received August 8, 2014

Report

Report Number
2649622-2014-08433
Event Type
Injury
Date Received
August 8, 2014
Date of Event
January 1, 2014
Report Date
June 4, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED PACING CAPTURE THRESHOLD IN THE RV (RIGHT VENTRICLE) WAS ELEVATED. THRESHOLD STABLE ~.75V, BEGINS RISE AROUND (B)(6) 2014. MAXIMUM OF 2.5V STARTING (B)(6) 2014.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT RISING THRESHOLDS, AND LOSS OF CAPTURE WERE OBSERVED ON THE RIGHT VENTRICULAR (RV) LEAD. IT WAS FURTHER REPORTED THAT RAPID BATTERY DEPLETION WAS SUSPECTED ON THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD). A PROCEDURE WAS SCHEDULED TO REPLACE THE RV LEAD AND ICD. THE LEAD AND DEVICE REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
469040 SPRINT QUATTRO SECURE DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 694765

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R DTBA2D1 ICD