ENRHYTHM MRI
Report
- Report Number
- 9614453-2014-01760
- Event Type
- Injury
- Date Received
- August 8, 2014
- Date of Event
- June 24, 2014
- Report Date
- June 24, 2014
- Manufacturer
- IPG MFG SWITZERLAND
- Product Code
- NVZ
- PMA / PMN Number
- P980035
- Removal / Correction Number
- Z-1440-2010
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- PHYSICIAN
Narratives
THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE U.S., HOWEVER, IT IS SIMILAR TO A DEVICE MARKETED IN THE U.S. THE EVENT OCCURRED OUTSIDE THE US AND IS BEING REPORTED DUE TO AN ALLEGED MALFUNCTION. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS REVEALED THE RETURNED DEVICE INDICATED HIGH IMPEDANCE IN THE BATTERY. (B)(4). THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION, BUT RETURNED PRODUCT TESTING FOUND THE DEVICE DID PERFORM AS DESCRIBED IN THE FIELD ACTION.
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) REACHED ELECTIVE REPLACEMENT INDICATOR (ERI) PREMATURELY DUE TO INCREASED BATTERY IMPEDANCE. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 469896 | ENRHYTHM MRI | PULSE GENERATOR, PERMANENT, IMPLANTABLE | NVZ | IPG MFG SWITZERLAND | EMDR01 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R |