FDA Adverse Event Malfunction Summary report: N

REVO MRI SURESCAN

MDR report key: 3990766 · Received August 8, 2014

Report

Report Number
3004209178-2014-14400
Event Type
Malfunction
Date Received
August 8, 2014
Date of Event
May 15, 2014
Report Date
May 15, 2014
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
LWP
PMA / PMN Number
P090013
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE ACTUAL PRODUCT WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA WAS COLLECTED FROM THE DEVICE AND WAS ANALYZED. ANALYSIS OF THE DEVICE MEMORY SHOWED THE BATTERY INDICATOR SIGNIFYING THAT IT THE TIME IS APPROACHING FOR DEVICE REPLACEMENT. CONCOMITANT PRODUCTS: (B)(4) LEAD, IMPLANTED 2011 (B)(6). (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS RECEIVED AND ANALYZED. THE RETURNED DEVICE INDICATED LOCK-UP IN AN INTEGRATED CIRCUIT. PERFORMANCE DATA WAS COLLECTED FROM THE DEVICE AND WAS ANALYZED. ANALYSIS OF THE DEVICE MEMORY SHOWED THE BATTERY INDICATOR SIGNIFYING THAT IT THE TIME IS APPROACHING FOR DEVICE REPLACEMENT. NOTCH BATTERY VOLTAGE TREND WAS OBSERVED AND ERI WAS NOTED TO BE TRIGGERED ON (B)(6) 2015.

Description of Event or Problem · 1

THE IMPLANTABLE PULSE GENERATOR (IPG) HAS REACHED END OF SERVICE (EOS). DEVICE HAS BEEN REMOVED AND REPLACED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DEVICE HAS UNEXPECTED LONGEVITY BUT HAD NOT REACHED ELECTIVE REPLACEMENT INDICATOR. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
469814 REVO MRI SURESCAN PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE LWP MEDTRONIC MED REL MEDTRONIC PUERTO RICO RVDR01

Patients

Seq Age Sex Outcome Treatment
1 00055 YR (B)(4) LEAD