FDA Adverse Event Injury Summary report: N

CONSULTA CRT-D

MDR report key: 3990765 · Received August 8, 2014

Report

Report Number
3004209178-2014-14399
Event Type
Injury
Date Received
August 8, 2014
Date of Event
May 21, 2014
Report Date
June 4, 2014
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
NIK
PMA / PMN Number
P010031
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THIS INFORMATION WAS RECEIVED FROM A COMPETITOR. ALL DATA PERTINENT TO THE EVENT IS PROVIDED. PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION. CONCOMITANT MEDICAL PRODUCTS: 1156T LEAD, IMPLANTED: (B)(6) 2010. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS REMOVED DUE TO PREMATURE BATTERY DEPLETION. IT WAS NOTED THAT THERE WERE CHRONIC HIGH RIGHT VENTRICULAR THRESHOLDS. THE ICD WAS REMOVED AND A NEW DEVICE WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
469556 CONSULTA CRT-D DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO NIK MEDTRONIC MED REL MEDTRONIC PUERTO RICO D224TRK

Patients

Seq Age Sex Outcome Treatment
1 00036 YR Hospitalization| R 694365V LEAD, 5076-52 LEAD