FDA Adverse Event Malfunction Summary report: N

ATTAIN PERFORMA

MDR report key: 3990761 · Received August 8, 2014

Report

Report Number
2649622-2014-08490
Event Type
Malfunction
Date Received
August 8, 2014
Date of Event
May 6, 2014
Report Date
May 6, 2014
Manufacturer
MPRI
Product Code
OJX
PMA / PMN Number
P080006
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES, HOWEVER, IT IS SIMILAR TO A DEVICE MARKETED IN THE U.S. THE EVENT IS BEING REPORTED DUE TO AN ALLEGED MALFUNCTION. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED, AND THE STYLET/GUIDEWIRE WAS KINKED/BUCKLED. THERE WAS BLOOD ON THE STY LET/GUIDEWIRE. THERE WAS BLOOD ON THE DISTAL CONDUCTOR OF THE LEAD AND IT WAS NOT OBSTRUCTED. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE THE PHYSICIAN WAS UNABLE TO REMOVE THE GUIDEWIRE FROM THE LEAD. THE LEAD DISLODGED AND WAS WITHDRAWN FROM THE PATIENT. THE GUIDEWIRE DID NOT ADVANCE OR WITHDRAW FROM THE LEAD WHEN PLACED ON THE SCRUB TABLE. A NEW LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
471299 ATTAIN PERFORMA DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE OJX MPRI 439888

Patients

Seq Age Sex Outcome Treatment
1 00070 YR GWR419588, GUIDEWIRE