SPRINT QUATTRO SECURE S

MDR report key: 3990711 · Received August 8, 2014

Report

Report Number
2649622-2014-08520
Event Type
Malfunction
Date Received
August 8, 2014
Date of Event
June 30, 2014
Report Date
June 30, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY. THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE ATTEMPTED SINGLE COIL RIGHT VENTRICULAR (RV) LEAD FAILED DEFIBRILLATION THRESHOLD TESTING. THE PATIENT WAS EXTERNALLY DEFIBRILLATED. THE LEAD WAS REMOVED AND REPLACED WITH A DUAL COIL LEAD WHICH PASSED THE DEFIBRILLATION THRESHOLD TEST AND REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
468196 SPRINT QUATTRO SECURE S DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 6935M55

Patients

Seq Age Sex Outcome Treatment
1 00056 YR