FDA Adverse Event Malfunction Summary report: N

SPRINT FIDELIS

MDR report key: 3990668 · Received August 8, 2014

Report

Report Number
2649622-2014-08532
Event Type
Malfunction
Date Received
August 8, 2014
Date of Event
May 9, 2014
Report Date
April 26, 2019
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: 429688 LEAD, IMPLANTED (B)(6) 2013; 5076-45 LEAD, IMPLANTED (B)(6) 2005. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE ACTUAL PRODUCT WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA WAS COLLECTED FROM THE DEVICE AND WAS ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED THE IMPEDANCE ON THE SUPERIOR VENA CAVA (SVC) DEFIBRILLATION COIL WAS BEYOND THE EXPECTED UPPER RANGE. MAX SVC DEFIB IMPEDANCE RISES FROM 69 OHM THE WEEK ENDING (B)(6) 2014 TO GREATER THAN 250 OHM THE WEEK ENDING (B)(6) 2014.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE LEAD WAS NOT RETURNED FOR ANALYSIS, HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED THE IMPEDANCE OF THE SUPERIOR VENA CAVA DEFIBRILLATION COIL WAS BEYOND THE EXPECTED UPPER RANGE. ANALYSIS OF THE DEVICE MEMORY INDICATED THE IMPEDANCE TREND OF THE SUPERIOR VENA CAVA DEFIBRILLATION COIL WAS VARIABLE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD MEASURED HIGH IMPEDANCE ON THE SUPERIOR VENA CAVA (SVC) COIL. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
467842 SPRINT FIDELIS DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 694858

Patients

Seq Age Sex Outcome Treatment
1 00066 YR D224TRK ICD