FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE S

MDR report key: 3990659 · Received August 8, 2014

Report

Report Number
2649622-2014-08551
Event Type
Injury
Date Received
August 8, 2014
Date of Event
June 11, 2014
Report Date
June 11, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES; HOWEVER, IT IS SAME/SIMILAR TO A DEVICE MARKETED IN THE U.S. THIS EVENT OCCURRED OUTSIDE THE US AND PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE LEAD WAS NOT RETURNED, ANALYSIS OF THE DEVICE MEMORY WAS PERFORMED AND NO ANOMALIES WERE FOUND.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED AND ANALYSIS WAS PERFORMED AND NO ANOMALIES WERE FOUND. THERE WERE NO ANOMALIES OBSERVED REGARDING THE RETURNED LEAD PROXIMAL SEGMENT. ALL ELECTRICAL TESTING WAS WITHIN SPECIFIED PARAMETERS. THE FULL LEAD WAS NOT RETURNED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT EXPERIENCED A SPONTANEOUS VENTRICULAR TACHYCARDIA (VT)/FAST - VENTRICULAR TACHYCARDIA (VT) EPISODE WHICH WAS APPROPRIATELY TREATED, BUT WITHOUT SUCCESS WITH ANTITACHYCARDIA PACING (ATP). THEN WITH THE RIGHT VENTRICULAR (RV) LEADFIRST SHOCK ACCELERATED IN VENTRICULAR FIBRILLATION (VF) WITH SIX SUCCESSIVE APPROPRIATE BUT UNSUCCESSFUL SHOCKS. THE PATIENT WAS REANIMATED/REVIVED AFTER SIX MINUTES. THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) AND THE RV LEAD WERE EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
467727 SPRINT QUATTRO SECURE S DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 693565

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| L| R D354TRG ICD