FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 3990620 · Received August 8, 2014

Report

Report Number
2649622-2014-08577
Event Type
Injury
Date Received
August 8, 2014
Date of Event
May 1, 2014
Report Date
June 27, 2014
Manufacturer
MPRI
Product Code
NVY
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MT, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4) ICD IMPLANTED: 2013 (B)(6); 429688 LEAD IMPLANTED: 2013 (B)(6). (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED T-WAVE OVERSENSING (TWOS). THE LEAD WAS REPROGRAMMED AND REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
472506 SPRINT QUATTRO SECURE DRUG-ELUTING PERMANENT DEFIBRILLATOR ELECTRODES NVY MPRI 6947M62

Patients

Seq Age Sex Outcome Treatment
1 00077 YR Required Intervention 407652 LEAD