FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 3990619 · Received August 8, 2014

Report

Report Number
2649622-2014-08579
Event Type
Injury
Date Received
August 8, 2014
Date of Event
May 8, 2014
Report Date
May 8, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT: D274DRG ICD IMPLANTED: 2010-(B)(6). (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: A PROXIMAL SEGMENT OF THE LEAD WAS RETURNED AND ANALYZED AND NO ANOMALIES WERE FOUND. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED APPARENT EXPLANT DAMAGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH IMPEDANCES AND SHORT INTERVAL COUNTS (SIC). THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
467881 SPRINT QUATTRO SECURE DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 6947

Patients

Seq Age Sex Outcome Treatment
1 00086 YR Hospitalization| R 5076-52 LEAD