Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT CONTINUED: 694965 LEAD, IMPLANTED (B)(6) 2007. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT RIGHT ATRIAL (RA) LEAD SHOWED OVERSENSING OF FAR FIELD R-WAVE (FFRW) AND SMALL P-WAVES. THE POST-VENTRICULAR ATRIAL BLANKING (PVAB) PERIOD HAD ALREADY BEEN PROGRAMMED. DECREASED SENSITIVITY WAS UNABLE TO RESOLVE THE ISSUE. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
467522 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5076-52

Patients

Seq Age Sex Outcome Treatment
1 00086 YR Required Intervention D314DRG ICD