FDA Adverse Event
Malfunction
Summary report: N
ATTAIN ABILITY
MDR report key: 3990551
·
Received August 8, 2014
Report
- Report Number
- 2649622-2014-08600
- Event Type
- Malfunction
- Date Received
- August 8, 2014
- Date of Event
- May 11, 2014
- Report Date
- May 11, 2014
- Manufacturer
- MPRI
- Product Code
- OJX
- PMA / PMN Number
- P080006
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYZED. NO ANOMALIES WERE FOUND. ELECTRICAL RESISTANCE AND CROSS CONTINUITY TEST RESULTS WERE WITHIN SPECIFICATION. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE, THE LEFT VENTRICULAR LEAD DEMONSTRATED HIGH THRESHOLDS AND THE PATIENT EXPERIENCED DIAPHRAGMATIC STIMULATION. THE LEAD WAS REMOVED AND A NEW LEAD WAS IMPLANTED SUCCESSFULLY. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 467142 | ATTAIN ABILITY | DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE | OJX | MPRI | 419688 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00068 YR |