FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3990547 · Received August 8, 2014

Report

Report Number
3004209178-2014-90865
Event Type
Malfunction
Date Received
August 8, 2014
Date of Event
July 4, 2014
Report Date
July 7, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED THE FUNCTIONAL TEST, INCLUDING THE DISPLACEMENT TEST, REWIND, BASIC OCCLUSION TEST, OCCLUSION TEST, PRIME/A33 TEST, AND THE EXCESSIVE NO DELIVERY TEST. THE INSULIN PUMP WAS PROGRAMMED WITH SEVERAL BOLUSES AND MONITORED. ALL BOLUSES DELIVERED PROPERLY AND WERE LISTED IN THE BOLUS HISTORY SCREEN. THE INSULIN PUMP CALCULATED THE ADDITIONAL BOLUS DELIVERIES AND WERE VERIFIED IN THE DAILY TOTAL SCREEN. THE INSULIN PUMP WAS THEN PROGRAMMED WITH MULTIPLE BASAL PROFILES AND MONITORED. ALL BASAL PROFILES DELIVERED THEIR INDICATED AMOUNTS PROPERLY AND WERE VERIFIED IN THE DAILY TOTAL SCREEN. NO DELIVERY ANOMALY, DAILY TOTAL ANOMALY, BASAL ANOMALY OR BOLUS ANOMALY NOTED. THE INSULIN PUMP HAD A CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

THE CUSTOMER'S FATHER REPORTED THAT THE CUSTOMER'S INSULIN PUMP WAS HAVING A DELIVERY ANOMALY AND THAT IT WAS UNDER DELIVERING. THE CUSTOMER'S BLOOD GLUCOSE WAS 120 MG/DL, THEN 115 MG/DL. DURING TROUBLESHOOTING, THE INSULIN PUMP PASSED THE SELF TEST, THE FIXED PRIME TEST, AND THE 5 UNITS BOLUS TEST. THE BASAL AND BOLUS SETTINGS WERE CHECKED AND WERE FOUND TO BE CORRECT. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
468275 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722WWS

Patients

Seq Age Sex Outcome Treatment
1