FDA Adverse Event Injury Summary report: N

ENRHYTHM DR

MDR report key: 3990539 · Received August 8, 2014

Report

Report Number
9614453-2014-01784
Event Type
Injury
Date Received
August 8, 2014
Date of Event
April 13, 2014
Report Date
May 16, 2014
Manufacturer
IPG MFG SWITZERLAND
Product Code
NVZ
PMA / PMN Number
P980035
Removal / Correction Number
Z-1440-2010
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION; HOWEVER, PERFORMANCE DATA WAS COLLECTED FROM THE DEVICE AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED A PARTIAL ELECTRICAL RESET. THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION.  BASED ON THE INFORMATION RECEIVED AND WITHOUT THE RETURN OF THE PRODUCT, IT COULD NOT DETERMINE THIS DEVICE PERFORMED AS DESCRIBED IN THE FIELD ACTION.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE HAD A RESET. THE DEVICE WAS INTERROGATED AND REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
467017 ENRHYTHM DR PULSE GENERATOR, PERMANENT, IMPLANTABLE NVZ IPG MFG SWITZERLAND P1501DR

Patients

Seq Age Sex Outcome Treatment
1 00089 YR Required Intervention