ENRHYTHM DR
Report
- Report Number
- 9614453-2014-01784
- Event Type
- Injury
- Date Received
- August 8, 2014
- Date of Event
- April 13, 2014
- Report Date
- May 16, 2014
- Manufacturer
- IPG MFG SWITZERLAND
- Product Code
- NVZ
- PMA / PMN Number
- P980035
- Removal / Correction Number
- Z-1440-2010
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT EVENT SUMMARY: THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION; HOWEVER, PERFORMANCE DATA WAS COLLECTED FROM THE DEVICE AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED A PARTIAL ELECTRICAL RESET. THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION. BASED ON THE INFORMATION RECEIVED AND WITHOUT THE RETURN OF THE PRODUCT, IT COULD NOT DETERMINE THIS DEVICE PERFORMED AS DESCRIBED IN THE FIELD ACTION.
THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).
IT WAS REPORTED THAT THE DEVICE HAD A RESET. THE DEVICE WAS INTERROGATED AND REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 467017 | ENRHYTHM DR | PULSE GENERATOR, PERMANENT, IMPLANTABLE | NVZ | IPG MFG SWITZERLAND | P1501DR |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00089 YR | Required Intervention |