FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3990412 · Received August 7, 2014

Report

Report Number
3004209178-2014-91059
Event Type
Malfunction
Date Received
August 7, 2014
Date of Event
July 4, 2014
Report Date
July 7, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

A CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, CRACKED DISPLAY WINDOW AND A CRACKED CASE NEAR DISPLAY WINDOW CORNERS WERE NOTED DURING THE VISUAL INSPECTION. ALL BUTTONS FUNCTIONED PROPERLY. NO BUTTON ERROR ALARMS WERE NOTED. HOWEVER, MOISTURE DAMAGE WAS NOTED ON KEYPAD TRACES.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

THE CUSTOMER REPORTED RECEIVING A BUTTON ERROR ALARM ON THE INSULIN PUMP. THE CUSTOMER'S BLOOD GLUCOSE WAS 160 MG/DL AT THE TIME THE ISSUE WAS REPORTED. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
463215 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAH A7723NAHJ

Patients

Seq Age Sex Outcome Treatment
1 15 YR