FDA Adverse Event Malfunction Summary report: N

530G INSULIN PUMP

MDR report key: 3990410 · Received August 7, 2014

Report

Report Number
3004209178-2014-91058
Event Type
Malfunction
Date Received
August 7, 2014
Date of Event
July 7, 2014
Report Date
July 7, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZO
PMA / PMN Number
120010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP DID FUNCTION PROPERLY DURING REWIND, BASIC OCCLUSION, OCCLUSION, PRIME, EXCESSIVE NO DELIVERY AND DISPLACEMENT TEST. NO MOTOR ERROR ALARM NOTED. THE MOTOR WAS TESTED OUTSIDE THE DEVICE AND PASSED MOTOR TEST. THE DEVICE WAS RECEIVED WITH MINOR SCRATCHED DISPLAY WINDOW, CRACKED CASE AT DISPLAY WINDOW CORNERS AND CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

THE CUSTOMER REPORTED RECEIVING A MOTOR ERROR ALARM WHILE ATTEMPTING TO DO A BOLUS DELIVERY. THE CUSTOMER'S BLOOD GLUCOSE WAS 171 MG/DL AT THE TIME THE ISSUE WAS REPORTED. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
464864 530G INSULIN PUMP OZO OZO MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-751NAS A2751NASJ

Patients

Seq Age Sex Outcome Treatment
1 62 YR