PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3990396 · Received August 7, 2014

Report

Report Number
3004209178-2014-91065
Event Type
Malfunction
Date Received
August 7, 2014
Date of Event
July 7, 2014
Report Date
July 7, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP ALARMED FOR A BUTTON ERROR DUE TO CORRODED KEYPAD TRACES. THE INSULIN PUMP HAD MINOR SCRATCHES ON THE DISPLAY WINDOW, A CRACKED DISPLAY WINDOW CORNER, CRACKED BATTERY TUBE THREADS, A CRACKED RESERVOIR TUBE WINDOW AND A MISSING END CAP STICKER.

Description of Event or Problem · 1

THE CUSTOMER REPORTED RECEIVING A BUTTON ERROR ALARM AFTER A BATTERY CHANGE. THE CUSTOMER DID NOT RECALL ANY EVENTS THAT MIGHT HAVE LED TO THE ALARM EXCEPT THAT IT WAS VERY HOT AND HUMID OUTSIDE. THE CUSTOMER'S BLOOD GLUCOSE WAS 148 MG/DL AT THE TIME THE ISSUE WAS REPORTED. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
464199 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-523NAS A4523NASJ

Patients

Seq Age Sex Outcome Treatment
1 51 YR