FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3990350 · Received August 7, 2014

Report

Report Number
3004209178-2014-91277
Event Type
Malfunction
Date Received
August 7, 2014
Date of Event
June 18, 2014
Report Date
July 9, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

A CRACKED RESERVOIR TUBE LIP, MINOR SCRATCHES ON THE DISPLAY WINDOW AND A CRACKED CASE NEAR THE DISPLAY WINDOW CORNERS WERE NOTED DURING THE VISUAL INSPECTION. THE INSULIN PUMP HAD INTERMITTENT BUTTON RESPOSE DUE TO A FLATTENED DOME SWITCH ON THE UP ARROW BUTTON.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THE UP ARROW BUTTON ON THE INSULIN PUMP IS NOT RESPONDING PROPERLY. CUSTOMER STATED THE BUTTON IS NOT CLICKING WHEN BEING PRESSED AND HAS TO PRESS IT MULTIPLE TIMES FOR IT TO WORK. CUSTOMER STATED HE DOES NOT RECALL ANY SIGNIFICANT EVENTS LEADING TO THE KEYPAD ISSUE. BLOOD GLUCOSE LEVEL IS 185 MG/DL. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
464484 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAL

Patients

Seq Age Sex Outcome Treatment
1 50 YR