PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2014-91275
- Event Type
- Malfunction
- Date Received
- August 7, 2014
- Date of Event
- July 9, 2014
- Report Date
- July 9, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX
- Reporter Occupation
- PATIENT
Narratives
INSULIN PUMP ALARMED MOTOR ERROR DURING BASIC OCCLUSION TEST DUE TO LOOSE AND FLUSH DRIVE SUPPORT DISK. NO PRIME ALARMS WERE NOTED. INSULIN PUMP RECEIVED WITH CRACKED CASE AT DISPLAY WINDOW CORNERS AND WINDOW DISPLAY. INSULIN PUMP RECEIVED WITH MINOR SCRATCHES ON LCD WINDOW.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
CUSTOMER REPORTED THAT SHE RECEIVED AN ALARM ON THE INSULIN PUMP DURING PRIME. CUSTOMER STATED SHE IS NOT CONNECTED TO THE INSULIN PUMP AND THAT IT WAS NOT DROPPED OR BUMPED. CUSTOMER STATED THE DRIVE SUPPORT CAP IS RECESSED. BLOOD GLUCOSE IS 117 MG/DL. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 463468 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-722NAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 29 YR |