PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3990334 · Received August 7, 2014

Report

Report Number
3004209178-2014-91283
Event Type
Malfunction
Date Received
August 7, 2014
Date of Event
July 9, 2014
Report Date
July 9, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT FAMILY MEMBER OR FRIEND

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

INSULIN PUMP WAS RECEIVED WITH A PRIME ALARM DURING BASIC OCCLUSION TEST AND UNABLE TO PRIME DURING PRIME TEST DUE TO PROTRUDED AND LOOSE DRIVE SUPPORT DISK. INSULIN PUMP HAD MINOR SCRATCHES ON LCD WINDOW, CRACKED DISPLAY WINDOW CORNER, CRACKED BATTERY TUBE THREADS, CRACKED RESERVOIR TUBE LIP AND CRACKED RESERVOIR TUBE WINDOW.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER RECEIVED AN ALARM ON THE INSULIN PUMP, INDICATING THE FORCE SENSOR DID NOT DETECT THE RESERVOIR DURING SEATING. THE PATIENT'S BLOOD GLUCOSE WAS 182 MG/DL. IT WAS ADVISED THAT THE CUSTOMER DISCONTINUE USE OF THE INSULIN PUMP AND REVERT TO A BACK-UP PLAN. NOTHING FURTHER WAS REPORTED.