FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 3990246 · Received August 7, 2014

Report

Report Number
3008973940-2014-00285
Event Type
Injury
Date Received
August 7, 2014
Date of Event
June 19, 2014
Report Date
June 20, 2014
Manufacturer
MEDTRONIC SINGAPORE OPERATIONS
Product Code
DTB
PMA / PMN Number
P930039
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND. THE DISTAL LOW VOLTAGE (LV) ELECTRODE OF THE LEAD WAS COVERED IN BLOOD. THE DISTAL LV ELECTRODE OF THE LEAD WAS COVERED IN BODY TISSUE/FIBROTIC GROWTH.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Additional Manufacturer Narrative · 1

CORRECTED INFORMATION: NO EVAL EXPLAIN CODE IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A FEW HOURS OF IMPLANTATION, THE LEAD HAD HIGH THRESHOLD AND HIGH IMPEDANCE. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
463205 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MEDTRONIC SINGAPORE OPERATIONS 5076-58

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R