FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE S

MDR report key: 3990227 · Received August 7, 2014

Report

Report Number
2649622-2014-08738
Event Type
Malfunction
Date Received
August 7, 2014
Date of Event
May 14, 2014
Report Date
May 14, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4). EVALUATION SUMMARY: ANALYSIS WAS PERFORMED AND NO ANOMALIES WERE FOUND; FULL LEAD RETURNED AND ANALYZED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING IMPLANT ATTEMPT, DFT (DEFIBRILLATION THRESHOLD) TESTING FAILED IN MULTIPLE VECTORS WITH HIGH DFTS NOTED, AND THE PATIENT COULD NOT BE DEFIBRILLATED. THE LEAD WAS NOT USED AND ANOTHER LEAD WAS SUCCESSFULLY IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
463532 SPRINT QUATTRO SECURE S DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 6935M62

Patients

Seq Age Sex Outcome Treatment
1 00055 YR