FDA Adverse Event Injury Summary report: N

CONCERTO II CRT-D

MDR report key: 3990216 · Received August 7, 2014

Report

Report Number
3004209178-2014-14463
Event Type
Injury
Date Received
August 7, 2014
Date of Event
May 21, 2014
Report Date
May 21, 2014
Manufacturer
MEDTRONIC PUERTO RICO
Product Code
NIK
PMA / PMN Number
P010031
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVALUATION: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION. CONCOMITANT MEDICAL PRODUCTS: PRODUCT ID: 6949-65, DEFIB LEAD, IMPLANTED: (B)(6) 2004; 4194-88, NON DEFIB LEAD, IMPLANTED: (B)(6) 2004. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE TRIGGERED THE ELECTIVE REPLACEMENT INDICATOR (ERI) AND PREMATURE BATTERY DEPLETION WAS SUSPECTED. THE DEVICE WAS EXPLANTED AND REPLACED. IT WAS ALSO REPORTED THAT THERE WAS HIGH THRESHOLDS ON THE LEFT VENTRICULAR (LV) LEAD. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
463632 CONCERTO II CRT-D DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO NIK MEDTRONIC PUERTO RICO D274TRK

Patients

Seq Age Sex Outcome Treatment
1 00062 YR Hospitalization| R 4592 NON DEFIB LEAD