ATTAIN ABILITY PLUS

MDR report key: 3990209 · Received August 7, 2014

Report

Report Number
2649622-2014-08752
Event Type
Malfunction
Date Received
August 7, 2014
Date of Event
May 19, 2014
Report Date
May 19, 2014
Manufacturer
MPRI
Product Code
OJX
PMA / PMN Number
P080006
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE THE LEAD WAS ATTEMPTED TO BE IMPLANTED BUT HAD PULLED BACK FROM THE DISTAL POSITION IN THE CS (CORONARY SINUS) LATERAL BRANCH RESULTING IN PHRENIC NERVE STIMULATION. THE LEAD WAS REMOVED AND A DIFFERENT LEAD WAS IMPLANTED INSTEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
463503 ATTAIN ABILITY PLUS DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE OJX MPRI 429688

Patients

Seq Age Sex Outcome Treatment
1 00050 YR