FDA Adverse Event Malfunction Summary report: N

PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE

MDR report key: 3990202 · Received August 7, 2014

Report

Report Number
2183613-2014-01047
Event Type
Malfunction
Date Received
August 7, 2014
Date of Event
March 20, 2014
Report Date
June 5, 2014
Manufacturer
MEDTRONIC MILACA, INC.
Product Code
LWP
PMA / PMN Number
P820003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: ANALYSIS CONFIRMED THE CUSTOMER COMMENT OF CASE DAMAGE, BOTH THE UPPER AND LOWER CASES ARE BROKEN. ANALYSIS ALSO FOUND THAT THE BATTERY RELEASE, LEAD FLEX COVER AND BATTERY DRAWER WERE CONTAMINATED, ONE SIDE BAIL COVER WAS BROKEN, THE RING COVER AND RING BAIL WERE MISSING, THE BATTERY CONTACTS WERE COMPRESSED AND THE KEYBOARD WAS SCRATCHED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE EXTERNAL PULSE GENERATOR (EPG) HAD CASE DAMAGE. THE EPG WAS RETURNED FOR SERVICE. THERE WAS NO PATIENT INVOLVEMENT. IT WAS FURTHER REPORTED THAT THE GENERATOR SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER¿S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
464499 PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE LWP MEDTRONIC MILACA, INC. 5388

Patients

Seq Age Sex Outcome Treatment
1