FDA Adverse Event Injury Summary report: N

CAPSUREFIX MRI SURESCAN

MDR report key: 3990197 · Received August 7, 2014

Report

Report Number
2649622-2014-08768
Event Type
Injury
Date Received
August 7, 2014
Date of Event
May 14, 2014
Report Date
May 14, 2014
Manufacturer
MPRI
Product Code
NVN
PMA / PMN Number
P090013
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: 5086MRI58 LEAD, IMPLANTED: (B)(6) 2012. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: (B)(4): THE FULL LEAD WAS RETURNED, ANALYZED AND THE OUTER INSULATION OF THE LEAD DEVELOPED A BREACH DUE TO A DEPRESSION WHILE IN VIVO.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) AND TWO LEADS WERE REMOVED AND REPLACED DUE TO OVERSENSING OF HIGH FREQUENCY NOISE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
463500 CAPSUREFIX MRI SURESCAN DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE NVN MPRI 5086MRI52

Patients

Seq Age Sex Outcome Treatment
1 00065 YR Hospitalization| R RVDR01 IPG