FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE S

MDR report key: 3990179 · Received August 7, 2014

Report

Report Number
2649622-2014-08747
Event Type
Injury
Date Received
August 7, 2014
Date of Event
May 23, 2014
Report Date
May 23, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYZED. PRIMARY ANALYSIS REVEALED NO ANOMALIES. HOWEVER, TTHE OUTER INSULATION OF THE LEAD WAS EXTRINSICALLY BREACHED DUE TO A CUT. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT. CONCOMITANT: OTHER PRODUCT: 419688 CRDM NON-DEFIB LEAD IMPLANTED: 2014-(B)(6); 407652 CRDM NON-DEFIB LEAD IMPLANTED: 2014-(B)(6). (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE, THE PHYSICIAN ACCIDENTALLY PUT A SUTURE THROUGH THE LEAD. THE LEAD WAS EXPLANTED AND A NEW LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
464428 SPRINT QUATTRO SECURE S DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 6935M62

Patients

Seq Age Sex Outcome Treatment
1 00066 YR Hospitalization| R DTBA1D4 BIV ICD