FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX MRI SURESCAN

MDR report key: 3990135 · Received August 7, 2014

Report

Report Number
2649622-2014-08807
Event Type
Malfunction
Date Received
August 7, 2014
Date of Event
May 20, 2014
Report Date
May 20, 2014
Manufacturer
MPRI
Product Code
NVN
PMA / PMN Number
P090013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE ATTEMPTED RIGHT VENTRICULAR (RV) LEAD WAS DIFFICULT TO FIXATE AND EXHIBITED SUBOPTIMAL SENSING FUNCTION. THE LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
465021 CAPSUREFIX MRI SURESCAN DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE NVN MPRI 5086MRI58

Patients

Seq Age Sex Outcome Treatment
1 00077 YR MCS-P3-29-AOA AORTIC VALVE