FDA Adverse Event Injury Summary report: N

PROTECTA XT CRT-D

MDR report key: 3990098 · Received August 7, 2014

Report

Report Number
9614453-2014-01818
Event Type
Injury
Date Received
August 7, 2014
Date of Event
May 13, 2014
Report Date
May 13, 2014
Manufacturer
IPG MFG SWITZERLAND
Product Code
NIK
PMA / PMN Number
P010031
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. THE DEVICE MET 88% OF EXPECTED LONGEVITY. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL. CONCOMITANT PRODUCTS: 1488TC ST. JUDE LEAD, IMPLANTED: (B)(6) 2001; 4193-78 LEAD, IMPLANTED: (B)(6) 2003. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE WAS SUSPECTED FOR A BATTERY LONGEVITY ISSUE. THE DEVICE WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
464965 PROTECTA XT CRT-D DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO NIK IPG MFG SWITZERLAND D314TRG

Patients

Seq Age Sex Outcome Treatment
1 00070 YR Hospitalization| R 1559 VENTRITEX LEAD