FDA Adverse Event Injury Summary report: N

REVEAL LINQ

MDR report key: 3990092 · Received August 7, 2014

Report

Report Number
9614453-2014-01817
Event Type
Injury
Date Received
August 7, 2014
Date of Event
June 16, 2014
Report Date
June 16, 2014
Manufacturer
IPG MFG SWITZERLAND
Product Code
DSI
PMA / PMN Number
K132649
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS OF THE DEVICE INDICATED SHORTING/LOW IMPEDANCE IN AN INTEGRATED CIRCUIT.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE IMPLANTABLE CARDIAC MONITOR (ICM) DEVICE INTERROGATED SUCCESSFULLY SOON AFTER IMPLANT, BUT ONE WEEK LATER IT DID NOT INTERROGATE EVEN WITH USE OF MULTIPLE PROGRAMMERS. THE DEVICE WAS EXPLANTED AND WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
465249 REVEAL LINQ DETECTOR AND ALARM, ARRHYTHMIA DSI IPG MFG SWITZERLAND LNQ11

Patients

Seq Age Sex Outcome Treatment
1 00037 YR Hospitalization| R