FDA Adverse Event Malfunction Summary report: N

CAPSURE SP NOVUS

MDR report key: 3990078 · Received August 7, 2014

Report

Report Number
2649622-2014-08833
Event Type
Malfunction
Date Received
August 7, 2014
Date of Event
May 28, 2014
Report Date
May 28, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P830061
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED THE IMPEDANCE ON THE ATRIAL PACING LEAD WAS BEYOND THE EXPECTED UPPER RANGE. AUTO LEAD DIAGNOSTICS SHOWS HIGH ATRIAL LEAD IMPEDANCE THROUGHOUT RECORD TO GREATER THAN 20,000 OHMS. ATRIAL LEAD WARNING OCCURRED ON (B)(6) 2013.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD WAS EXHIBITING HIGH, OUT OF RANGE IMPEDANCE. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
462887 CAPSURE SP NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 4592-53

Patients

Seq Age Sex Outcome Treatment
1 KDR703 IPG