FDA Adverse Event Injury Summary report: N

VIRTUOSO VR

MDR report key: 3990041 · Received August 7, 2014

Report

Report Number
9614453-2014-01822
Event Type
Injury
Date Received
August 7, 2014
Date of Event
February 10, 2014
Report Date
May 10, 2014
Manufacturer
IPG MFG SWITZERLAND
Product Code
LWS
PMA / PMN Number
P980016
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. A HIGH CURRENT DRAIN CONDITION WAS FOUND DURING DEVICE ANALYSIS. ELECTRICAL ANALYSIS REVEALED THE CAUSE OF THE HIGH CURRENT DRAIN WAS CURRENT LEAKAGE IN THE BATTERY FILTER CAPACITORS. THE DEVICE MET 53% OF THE EXPECTED LONGEVITY. (B)(4).

Description of Event or Problem · 1

THE DEVICE WAS RETURNED TO THE MANUFACTURER, ANALYZED, TESTED OUT OF SPECIFICATION AND IS BEING REPORTED BASED ON ANALYSIS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS LATER REPORTED THAT THE DEVICE HAD TRIGGERED RECOMMENDED REPLACEMENT TIME (RRT) AND WAS ELECTIVELY REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
462566 VIRTUOSO VR DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS IPG MFG SWITZERLAND D164VWC

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R