FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 3990017 · Received August 7, 2014

Report

Report Number
2649622-2014-08852
Event Type
Injury
Date Received
August 7, 2014
Date of Event
June 15, 2014
Report Date
June 15, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: MODEL: 407645, LEAD; IMPLANT: (B)(6) 2014. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED. ANALYSIS WAS PERFORMED AND NO ANOMALIES WERE FOUND.

Description of Event or Problem · 1

IT WAS REPORTED THAT TWO DAYS POST IMPLANT, A PATIENTS RIGHT VENTRICULAR (RV) LEAD EXHIBITED INCREASED THRESHOLDS AND LOSS OF CAPTURE. AN X-RAY REVEALED AN INFERIOR WALL PERFORATION. THE LEAD WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
462485 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5076-52

Patients

Seq Age Sex Outcome Treatment
1 00060 YR Hospitalization| L| R ADDRL1, IPG