HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 3985265 · Received August 6, 2014

Report

Report Number
1416980-2014-25535
Event Type
Malfunction
Date Received
August 6, 2014
Date of Event
July 3, 2014
Report Date
July 12, 2014
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

COMPLAINT NO: CMPLNT-(B)(4). DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, AN ALARM INDICATIVE OF A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE WAS IDENTIFIED. AS THE CASSETTE WAS NOT RETURNED, A DEVICE ANALYSIS CANNOT BE COMPLETED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, A SYSTEM ERROR 2240 ALARM (AIR IN SET/LINE) WAS IDENTIFIED IN THE LOG WHICH INDICATES A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE. THE ALARM OCCURRED ON (B)(6) 2014 00:49:44. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
460084 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 89 YR HOMECHOICE