FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3984628 · Received August 6, 2014

Report

Report Number
3004209178-2014-90615
Event Type
Malfunction
Date Received
August 6, 2014
Date of Event
July 4, 2014
Report Date
July 7, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK
Reporter Occupation
PATIENT FAMILY MEMBER OR FRIEND

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

NO BUTTON ERROR ALARM WAS NOTED. THE INSULIN PUMP HAD NO BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. THE INSULIN PUMP HAD MINOR SCRATCHES ON THE DISPLAY WINDOW, A CRACKED CASE AT DISPLAY WINDOW CORNERS, A SCRATCHED RESERVOIR TUBE WINDOW, A CRACKED RESERVOIR TUBE AND A CRACKED RESERVOIR TUBE.

Description of Event or Problem · 1

CUSTOMER'S MOTHER REPORTED THAT THEY RECEIVED A BUTTON ERROR ALARM ON THE INSULIN PUMP. SHE STATED THAT CHANGING THE BATTERY DID NOT RESOLVE THE ISSUE AND THEY WERE UNABLE TO CLEAR THE ALARM. THE BLOOD GLUCOSE READING WAS NOT KNOWN. SHE WAS UNSURE OF WHAT LED UP TO THESE ISSUES. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
460020 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 23 YR