FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3984607 · Received August 6, 2014

Report

Report Number
2531779-2014-22592
Event Type
Malfunction
Date Received
August 6, 2014
Report Date
July 27, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 09/22/2014 WITH THE FOLLOWING FINDINGS: THE PUMP POWERS ON WITH AUDIBLE AND VIBRATORY SOUNDS. A REWIND, LOAD AND PRIME SEQUENCE WERE PERFORMED WITH NO ISSUES. PUMP WAS EXERCISED FOR A 24HR DURATION PERIOD AND NO SELF SUSPENDING OCCURRED DURING THIS TIME PERIOD. NO HYPERSENSITIVE KEYS WERE OBSERVED ON KEYPAD. PUMP WAS ABLE TO BE MANUALLY SUSPENDED AND MANUALLY RESUMED WITH NO ISSUES. UNABLE TO DUPLICATE THE COMPLAINT. THE DISPLAY SCREEN HAS A PINKISH CONTRAST. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A HISTORY/SETTINGS (SUSPEND) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
460014 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 7 YR