530G INSULIN PUMP
Report
- Report Number
- 3004209178-2014-90612
- Event Type
- Malfunction
- Date Received
- August 6, 2014
- Date of Event
- July 2, 2014
- Report Date
- July 7, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZO
- PMA / PMN Number
- 120010
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.
CUSTOMER REPORTED RECEIVING AN ALARM FOR A REBOOT AND A MOTOR ERROR ALARM ON THE INSULIN PUMP. SHE COULD NOT RECALL ANY SIGNIFICANT EVENT LEADING TO THIS ALARM AND STATED THAT THE INSULIN PUMP HAD NOT BEEN EXPOSED TO A STRONG MAGNETIC FIELD OR MRI. CUSTOMER WAS ASSISTED IN CLEARING THE ALARM AND STATED THAT SHE WAS ABLE TO REWIND THE INSULIN PUMP. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 459230 | 530G INSULIN PUMP | OZO | OZO | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-751NAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 35 YR |