PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2014-90584
- Event Type
- Malfunction
- Date Received
- August 6, 2014
- Date of Event
- July 6, 2014
- Report Date
- July 6, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.
FAILURE ANALYSIS REVEALED THAT NO BUTTON ERROR ALARM NOTED. HOWEVER, THE ACT AND UP ARROW BUTTONS DID NOT RESPOND DUE TO CORROSION ON KEYPAD TRACES. THE DEVICE WAS RECEIVED WITH CRACKED DISPLAY WINDOW, CRACKED DISPLAY WINDOW CORNERS, CRACKED BELT CLIP SLOT, CRACKED RESERVOIR TUBE LIP, AND MISSING END CAP STICKER.THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.
IT WAS REPORTED THAT THE INSULIN PUMP HAD AN UNRESPONSIVE KEYPAD. ALSO, IT WAS REPORTED THAT THE INSULIN PUMP ALARMED BUTTON ERROR THREE MONTHS PRIOR. THE CUSTOMER'S BLOOD GLUCOSE WAS 12.2 MMOL/L. THE DISCONTINUATION OF THE INSULIN PUMP WAS ADVISED. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 459602 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-754WWL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |